FRONT OF DEVICEComponents of the NOT014224-2 System REAR VIEW - SIDE VIEWVS Ultra™ & VS Integra™COMPONENTSIllustrations ACCESSORIESView from above
5Description of the deviceThe circuit connection area comprises:• An air outlet• Single circuit support with exhalation and pressure line connection.F
6Figure 5: Detailed view of connection for a double circuit.Rear view On the rear of the device, take particular note of the location of the following
7Description of the deviceFigure 6: Rear view of device.Side viewSome device models are fitted with a metal connector for attaching an oxygen supply (
82.3 Circuit Depending on the ventilation mode prescribed by your physician, you will be using one of the following: • Single circuit• Single circuit
9Description of the deviceExternal battery (optional)An external battery increases the utilisation period away from a mains power supply, and also pro
102.5 Device control panelThe device control panel comprises an LCD screen and a keypad.Figure 10: Front view.The buttons on the keypad are listed in
11Connection procedures3 Connection proceduresPlace the device on a flat surface. Ensure the area is dust-free, and cleared of any objects that could
12Notes:• If your device is fitted with a rechargeable built-in battery (available as an optional extra with the VS Integra), it can be used without a
13Connection proceduresOption 1 – Single circuit1. Connect one end of the horizontally-ribbed tube firmly to the air outlet, as shown in the Figure be
14Figure 13: Connection of single circuit with expiratory valve and pressure line. See the next Figure for a detailed view of the connection panel. Fi
CONNECTING THE ACCESSORIESOption 1 - Single circuit Option 3 - Double circuitExternal batteryPlug for supplemental oxygenOption 2 - Single circuit wit
15Connection proceduresNow that your circuit is connected to the device, you can connect the mask to the other end of the circuit.Option 3 – Double ci
163.4 Example of fully-assembled systemFigure 16: Example of fully-assembled system (single circuit with expiratory valve and pressure line).If your s
17Connection proceduresThis Figure is a detailed view of the connection of the cable to the device .Figure 18: Detailed view of connection to the devi
18Connecting supplemental oxygen (optional)The instructions in this section apply only to the connection of supplemental oxygen to your device. Refer
19Switching on and shutting down4 Switching on and shutting down4.1 StandbyWhen your device is connected to the mains power supply, it goes into stand
21Cleaning and Maintenance5 Cleaning and MaintenanceWe recommend that you maintain your device and accessories regularly.!CAUTIONIf you use the follow
22!WARNINGTo avoid any risk of electric shock, never immerse the device or power cable in water. Always unplug the device before cleaning and be sure
23Troubleshooting6 TroubleshootingAlarms will alert you to any problems with your device. The instructions in this section will help you to identify t
24Low pressure alarm (LP)This alarm also indicates a wrongly connected or disconnected circuit.Reconnect the circuit.High pressure alarm (HP)A compone
VS Ultra™ & VS Integra™Patient manualEnglishNOT014224-2 07 10
25Technical Specifications7 Technical SpecificationsCE-marked device: 7.1 DimensionsL: LengthH: HeightD: Depth7.2 WeightVS Integra model (red) without
267.3 Power supplyMains power230/100 VAC; 50-60 Hz; 52 VA, 630 mA.External power supply26 VDC; 3 A maximum.Built-in battery (option for VS Integra)NiM
27Appendix8 AppendixTravelling with your ventilatorFor long journeys it is advisable to carry your device in its travel bag, with the following access
28
29IndexIndexAaccessories 8, 16air outlet 5, 15alarm silence button 10, 19Bbattery 12battery holder 9, 16battery life 26built-in battery 12, 19Ccircuit
Global leaders in sleep and respiratory medicine www.resmed.comHME provider references*N0T014224*
ResMed Schweiz AG Viaduktstrasse Basel, Switzerland, +41 61 564 70 00 or 0800 00 2500 (toll free). Saime SAS (Manufacturer), Savigny-le-Temple, France
iiiContents1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 Definitions 11.2 User/Owner Responsibi
1Introduction1 IntroductionThis manual is provided with the device you have received, a medical ventilator in the VS range. It does not in any way rep
2The VS Integra and VS Ultra can both be used with single circuit, but only the VS Ultra model can be used with double circuit. Warnings• This manual
3Description of the device2 Description of the device2.1 Components The picture below shows the set of components available from your HME provider:• Y
42.2 Your device Front viewThe device comprises:• A control panel • An area for connecting the circuit .Figure 2: Front view of the device.The contro